Real-time monitoring of cleanroom laboratories using low cost sensors

Cited 0 time in wos
Cited 0 time in scopus
Title
Real-time monitoring of cleanroom laboratories using low cost sensors
Other Titles
저비용 센서를 이용한 청정 실험실 실시간 감시
Authors
Han, Yeongcheol
Jung, HyeJin
Moon, Jangil
Baek, Jongmin
Han, Changhee
Hur, Soon Do
Subject
Geology
Keywords
Arduinocleanroomfine dustair filterparticle counter
Issue Date
2019-02
Citation
Han, Yeongcheol, et al. 2019. "Real-time monitoring of cleanroom laboratories using low cost sensors". Journal of the Geological Society of Korea, 55(1): 141-148.
Abstract
An air filtration system (AFS) is necessary for laboratories that analyze trace components in natural or artificial samples liable to be contaminated by airborne particulates. In general, an AFS employs multi-stage air filters with a HEPA (high efficiency particulate air) or an ULPA (ultra-low penetration air) filter as the final filter to supply particle-free air to a laboratory. However, using these filters does not guarantee the desired cleanliness of air, and routine assessment of the particle levels in laboratory air is required especially given the potential increase in atmospheric fine particles in Korea. In this context, we have developed a real-time particle monitor (PaMo) using a low cost sensor module and a web based data logger. A temperature and humidity sensor was also included in the PaMo. Two PaMo units have been installed in two cleanroom laboratories for ice core research at the Korea Polar Research Institute. The number of particles larger than 0.3 μm, temperature and relative humidity were monitored from 29th March to 24th June 2018. This real-time monitoring proved that, although the particle concentrations in the room air increased with that in the outdoor air, the AFSs were able to achieve and maintain the cleanliness required for each laboratory. The PaMo will be able to identify particle sources other than the outdoor air, provide a guideline for when to replace filters and notify unexpected AFS failure. We suggest that the PaMo is an easy and effective alternative to expensive particle counters for laboratories that operate AFS.
URI
https://repository.kopri.re.kr/handle/201206/10823
DOI
http://dx.doi.org/10.14770/jgsk.2019.55.1.141
Type
Article
Station
해당사항없음
Indexed
KCI등재
Appears in Collections  
2018-2018, Reconstruction of past climate and environmental changes using high resolution ice core records in Victoria Land, Antarctica (18-18) / Hur, Soon Do (PE18040)
2017-2018, Reconstruction of past climate and environmental changes using high resolution ice core records in Victoria Land, Antarctica (17-18) / Hur, Soon Do (PE17040; PE18040)
Files in This Item

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse